Cardiff University | Prifysgol Caerdydd ORCA
Online Research @ Cardiff 
WelshClear Cookie - decide language by browser settings

Control of encrustation and blockage of Foley catheters

Stickler, David James, Jones, G. L. and Russell, Allan Denver 2003. Control of encrustation and blockage of Foley catheters. Lancet 361 (9367) , pp. 1435-1437. 10.1016/S0140-6736(03)13104-2

Full text not available from this repository.

Abstract

Urinary catheters often become encrusted and blocked by crystalline Proteus mirabilis biofilms. Results of experiments in a laboratory model of a Foley catheterised bladder infected with P mirabilis showed that when retention balloons were inflated with a solution of triclosan (10 g/L), the catheters drained freely for at least 7 days. Triclosan became impregnated throughout the silicone catheter material and completely inhibited the formation of crystalline biofilm, whereas catheters inflated with water became blocked in 24 h. Our observations suggest a way to control a common complication in patients with long-term indwelling bladder catheters.

Item Type: Article
Date Type: Publication
Status: Published
Schools: Biosciences
ISSN: 0140-6736
Last Modified: 24 Mar 2017 02:33
URI: https://orca.cardiff.ac.uk/id/eprint/1073

Citation Data

Cited 70 times in Scopus. View in Scopus. Powered By Scopus® Data

Actions (repository staff only)

Edit Item Edit Item