Cardiff University | Prifysgol Caerdydd ORCA
Online Research @ Cardiff 
WelshClear Cookie - decide language by browser settings

Variation in electrosurgical vessel seal quality along the length of a porcine carotid artery

Wyatt, Hayley Louise, Richards, Rosie, Pullin, Rhys ORCID: https://orcid.org/0000-0002-2853-6099, Yang, T. H. Jimmy, Blain, Emma Jane ORCID: https://orcid.org/0000-0001-8944-4254 and Evans, Samuel ORCID: https://orcid.org/0000-0003-3664-2569 2016. Variation in electrosurgical vessel seal quality along the length of a porcine carotid artery. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine 230 (3) , pp. 169-174. 10.1177/0954411915621092

[thumbnail of wyatt et al 2016.pdf]
Preview
PDF - Published Version
Available under License Creative Commons Attribution.

Download (1MB) | Preview

Abstract

Electrosurgical vessel sealing has been demonstrated to have benefits for both patients and practitioners, but significant variation in the strength of the seal continues to be a concern. This study aims to examine the variation in electrosurgical seal quality along the length of a porcine common carotid artery and explore the relationships between seal quality, vessel size and morphology. Additionally, the study aimed to investigate the minimum safety threshold for successful seals and the influence of vessel characteristics on meeting this requirement. A total of 35 porcine carotid arteries were sealed using the PlasmaKinetic Open Seal device (Gyrus). Each seal was burst pressure tested and a sample taken for staining with elastin van Gieson’s stain, with morphological quantification using image processing software ImageJ. With increasing distance from the bifurcation, there was an increase in seal strength and a reduction in both elastin content and vessel outer diameter. A significant correlation was found between burst pressure with both outer diameter (p<0.0001) and elastin content (p = 0.001). When considering the safe limits of operation, vessels of less than 5 mm in outer diameter were shown to consistently produce a seal of a sufficient strength (burst pressure > 360mmHg) irrespective of vessel morphology.

Item Type: Article
Date Type: Publication
Status: Published
Schools: Biosciences
Engineering
Subjects: T Technology > T Technology (General)
T Technology > TA Engineering (General). Civil engineering (General)
T Technology > TJ Mechanical engineering and machinery
Publisher: Sage
ISSN: 0954-4119
Funders: EPSRC
Date of First Compliant Deposit: 30 March 2016
Date of Acceptance: 13 November 2015
Last Modified: 09 May 2023 01:48
URI: https://orca.cardiff.ac.uk/id/eprint/85913

Citation Data

Cited 9 times in Scopus. View in Scopus. Powered By Scopus® Data

Actions (repository staff only)

Edit Item Edit Item

Downloads

Downloads per month over past year

View more statistics